Chotest Laser Interferometer Measurement System SJ6000



  • Item No.: SJ6000
  • Product name: SJ6000 Laser Interferometer Measurement System
  • Linear measuring range: 0~80 m
  • Resolution: 1 nm
  • Laser frequency accuracy: 0.05 ppm
  • Operating temperature: (0-40)℃
  • Dynamic capture rate: 50 kHz
  • Manufacturer: Chotest Technology Inc.
  • Country of origin: China



Technical data and information

System parameters:
     1. Measuring method: single frequency
     2. Laser frequency accuracy: 0.05ppm
     3. Dynamic capture rate: 50kHz
     4. Warm-up time: about 6 min
     5. Operating temperature: (0-40)
     6. Environment temperature: (0-40), humidity: 0-95%
     7. Storage temperature: -20-70
Environmental sensors:
     1. Atmospheric temperature sensor : ±0.1 (0-40), resolution: 0.01
     2. Material temperature sensor: ±0.1 (0-40), resolution: 0.01
     3. Atmospheric humidity sensor: ±5% (0-95%)
     4. Atmospheric pressure sensor: ±0.1kPa (65115)kPa
Linear measurement:
     1. Measuring range: (0-80)m
     2. Measuring accuracy: 0.5ppm (0-40)
     3. Measuring resolution: 1nm
     4. Maximum measuring speed: 4m/s
Angle measurement:
     1. Axial range: (0-15)m
     2. Measuring range: ±10°
     3. Accuracy: ±(0.02%R+0.1+0.24M)″ (R is indicating value, unit: ″; M is measured length in m)
     4. Measuring resolution: 0.1″
Flatness measurement:
     1. Axial range: (0-15) m
     2. Flatness measuring range: ±1.5 mm
     3. Accuracy: ±(0.2%R+0.02M2) μm (R is indicating value in μm; M is measured length in m)
     4. Substrate size: 180mm adjustable, 360mm adjustable
     5. Measuring resolution: 0.1μm
Straightness measurement:
Item Axis range Measuring range Accuracy Resolution
Short straightness (0.1-4)m ±3.0mm ±(0.5+0.25%R+0.15M2)μm 0.01μm
Long straightness (1-20)m ±3.0mm ±(5.0+2.5%R+0.015M2)μm 0.1μm
    Note: R is indicating value in μm; M is measured length in meters
Perpendicularity measurement:
Item Axis range Measuring range Accuracy Resolution
Short perpendicularity (0.1-3)m ±3.0mm ±(2.5+0.25%R+0.8M)μm/m 0.01μm
Long perpendicularity (1-15)m ±3.0mm ±(2.5+2.5%R+0.08M)μm/m 0.01μm
    Note: R is indicating value in μm; M is measured length in meters
Rotary axis measurement:
     1. Measuring range of angle: 0-360°
     2. Max axis rotation speed: 10rpm
     3. Pitch accuracy of precision turntable: ±1″
     4. Resolution: 0.1″


Подробно Product descriptions

Laser interferometer is recognized as a high precision, high sensitive measuring method by applying light wavelength as criterion, and is widely used in high-end manufacturing domain.

SJ6000 laser interferometer insists of high-frequency Helium-Neon laser generator from a USA supplier, high-precision environmental compensation modules, high-precision laser interference signal processing system, high-performance computer control system. By applying with thermal frequency stabilization technology of laser dual-longitudinal mode and geometric parameters interference optical path design, SJ6000 can output long-term stable and high-precision (0.05 ppm) laser quickly (about 6 minutes) which has powerful anti-interference performance. With different prism modules, it can measure linearity, angle, straightness, Flatness and perpendicularity, in addition it can analyze dynamic characteristics.